A blood-like control based on lyophilized hemoglobin with total Hb (tHb) and fractions of Hb-derivatives FO2Hb, FCOHb and FMetHb. Multiple levels are available covering the clinically relevant ranges to verify the performance of CO-OXimeters. This material is ideal for Proficiency Testing (PT).
Description
Intended Use
HemoxiTrol is an assayed real hemoglobin control intended for professional use in the verification of the precision and accuracy of CO-oximeters. The purpose of this quality control is to check wavelength and absorbance calibration and other performance related characteristics when measuring total Hb (tHb) and fractions of Hb-derivatives FO2Hb, FCOHb and FMetHb.
Analytes
ARTICLE CODE
069.001.001
069.002.001
069.003.001
069.004.001
069.005.001
069.006.001
Level 1
Level 2
Level 3
Level 4
Level 5
Level 6
pH
7.000 - 7.200
7.000 - 7.200
7.000 - 7.200
7.000 - 7.200
7.000 - 7.200
7.000 - 7.200
1
tHb
17.3 - 18.7
15.2 - 16.6
13.1 - 14.5
11.3 - 12.7
9.4 - 10.8
7.3 - 8.7
g/dL
O2Hb
90.0 - 100.0
80.0 - 90.0
70.0 - 80.0
60.0 - 70.0
50.0 - 60.0
40.0 - 50.0
%
COHb
0.0 - 5.0
0.0 - 10.0
5.0 - 15.0
10.0 - 20.0
15.0 - 25.0
25.0 - 35.0
%
MetHb
0.0 - 10.0
5.0 - 15.0
10.0 - 20.0
15.0 - 25.0
20.0 - 30.0
25.0 - 35.0
%
HHb
0.0 - 2.0
0.0 - 2.0
0.0 - 2.0
0.0 - 2.0
0.0 - 2.0
0.0 - 2.0
%
These are target ranges for manufacturing of the product. Individual lots have varying means and ranges due to a combination of factors, including assay method, instrument platform, reagent lot, matrix effects, etc. For accurate quality assessment and verification lot- specific values and ranges established by peer group results from CueSee® Compare should be used.
Details
Glass Vial
0.5 mL
1 per Box
Unopened
-24 - -16 °C
Opened
2 - 8 °C
Unopened
1095 days
Opened
N/A
Tutorials