Description

Validate the precision and accuracy of combined blood gas and electrolyte analyzers. Eurotrol GAS-ISE provides three physiological relevant levels simulating clinically significant ranges of acid/base and electrolyte balance, respiratory function, glucose, lactate and urea concentration.

Intended Use

Eurotrol GAS-ISE is an assayed quality control material intended for professional use in the verification of the precision and accuracy of combined blood gas and electrolyte analyzers. Eurotrol GAS-ISE is also developed to verify calibration, operating temperature and other performance characteristics.

Analytes

ARTICLE CODE
013.001.010
013.002.010
013.003.010
Level 1
Level 2
Level 3
pH
7.150 - 7.250
7.350 - 7.450
7.550 - 7.650
1
pO2
67.0 - 77.0
96.0 - 110.0
129.0 - 147.0
mmHg
pCO2
53.0 - 63.0
35.0 - 45.0
16.0 - 24.0
mmHg
Na+
117 - 123
137 - 143
157 - 163
mmol/L
K+
2.7 - 3.3
4.2 - 4.8
6.7 - 7.3
mmol/L
Li+
0.5 - 0.6
0.8 - 0.8
1.1 - 1.2
mmol/L
Ca2+
1.75 - 1.95
1.20 - 1.40
0.65 - 0.85
mmol/L
tCa
2.47 - 2.57
1.67 - 1.77
0.63 - 0.73
mmol/L
tMg
0.40 - 0.60
0.80 - 1.00
1.40 - 1.60
mmol/L
Cl-
77 - 83
97 - 103
117 - 123
mmol/L
Glu
2.9 - 3.1
5.9 - 6.1
12.4 - 12.6
mmol/L
Lac
7.20 - 7.80
2.70 - 3.30
0.90 - 1.50
mmol/L
Urea
0.5 - 1.1
3.0 - 3.7
7.2 - 7.8
mmol/L
These are target ranges for manufacturing of the product. Individual lots have varying means and ranges due to a combination of factors, including assay method, instrument platform, reagent lot, matrix effects, etc. For accurate quality assessment and verification lot- specific values and ranges established by peer group results from CueSee® Compare should be used.

Details

Ampoule
2.5 ± 0.1 mL
10 per Box
Unopened 2 - 8 °C
Opened 2 - 8 °C
Unopened 365 days
Opened 30 sec

Documents

Tutorials

CueSee® Online